INLINE PROCESS VISCOMETER
The Brookfield FAST system has a rugged design with non-moving parts. It has been designed to perform day after day under the extreme demands in printing, coating and adhesive operations - with no maintenance. Utilizing a high frequency, micro-rotational core, which is unaffected by press vibration, this worry-free system can measure and control the viscosity of your inks, coatings, and adhesives from start to finish.allowing the operators to handle other press concerns.
FEATURES & BENEFITS
- Unique micro rotational sensor
- No moving parts, no maintenance
- Unaffected by external vibration
- Flexible vertical or horizontal installation
- True temperature readings to 200°C
- Maintains correct color throughout the process
- Compact: only 8.5 inches tall
- Worldwide support
FAST/MXTS SENSOR SPECIFICATIONS
Measurement Type | Torsionally oscillating probe |
Measurement Range | 1 to 3,000 (optional 12,000) cSt using an NIST traceable, Newtonian fluid measured at 25°C |
Process Connections | ¾” standard, (1, 2½” tri-clamp or 3A design, optional) |
Repeatability | ±1.0% of reading |
Wetted Surfaces | 316L stainless steel Isolast for probe Viton for chamber housing |
Sensor O-Ring: | (EDPM or Kalrez®, optional) |
Temperature (Fluid) | -4°F to 392°F (-20°C to 200°C) |
Pressure Range | Vacuum to 200 pig max., 400 psig option |
FMXTS TRANSMITTER SPECIFICATIONS
Viscosity Range (cSt) (field selectable) | 0 to 10, 0 to 50, 0 to 100, 0 to 250, 0 to 500, 0 to 1000, 0 to 1500, 0 to 2000, 0 to 12500 |
Analog Output (2) | 4-20 mA (non-isolated) CH1 = viscosity CH2 = temperature |
Serial Port (2) | Port 1 = RS232, simple, read only Port 2 = ½ Duplex RS485, Modbus RTU driver |
Electronics Packaging | Wall mount, NEMA-4 (IP65), 8” x 8” x 6” (203 x 203 x 152mm) 32°F to 104°F (0°C to 40°C) |
Interconnecting Cable | 25 ft (7.6m) standard Optional: 50 ft (15.2m), 75 ft (22.8m) and 100 ft (30.4m) |
Supply Voltage | 18-30VDC or 115VAC/60 Hz or 230VAC/50 Hz (±10% max), power draw <1 amp |