UV-2600/2700

Hardware Specifications

ItemUV-2600UV-2700
Photometric systemDouble-beam optics
Photometric systemCzerny-Turner mounting Single monochromator Lo-Ray-Ligh grade blazed holographic gratingCzerny-Turner mounting Double monochromator Lo-Ray-Ligh grade blazed holographic grating
DetectorR-928  Photomultiplier
Light source50 W halogen lamp, deuterium lamp, light source auto position adjustment built in
Setting wavelength range185–1400 nm
Measurement wavelength range185–900 nm
220 nm to 1400 nm when the ISR-2600Plus Integrating Sphere Attachment is used.The ISR-2600Plus Integrating Sphere Attachment cannot be used
Wavelength accuracy± 0.1 nm (656.1 nm D2), ± 0.3 nm (all range)
Wavelength repeatability± 0.05 nm
Wavelength scanning speedWavelength slew rate: about 14000 nm/min Wavelength scan rate: about 4000 to 0.5 nm/min
Wavelength settingAt 1 nm units for scan start and scan end wavelengths, and 0.1 nm units for other wavelengths
Lamp interchange wavelengthAuto switching synchronized with wavelength; switching range selectable between 290 and 370 nm (0.1 nm units)
Spectral bandwidth0.1/0.2/0.5/1/2/5 nm L2/L5 (low stray-light mode)
Resolution0.1 nm
Stray lightMax. 0.005 % (220 nm, NaI) Max. 0.005 % (340, 370 nm, NaNO2) Max. 1 % (198 nm, KCl)Max. 0.00005 % (220 nm, NaI) Max. 0.00002 % (340, 370 nm, NaNO2) Max. 1 % (198 nm, KCl)
Photometric modesAbsorbance (Abs.), transmittance (%), reflectance (%), energy €
Photometric rangeAbsorbance: −5 to 5 Abs Transmittance, reflectance: 0 to 100000 %Absorbance: −8.5 to 8.5 Abs Transmittance, reflectance: 0 to 100000 %
Photometric accuracy± 0.002 Abs (0.5 Abs) ± 0.003 Abs (1 Abs) ± 0.006 Abs (2 Abs) ± 0.3 %T Measured using NIST930D/NIST1930 or equivalent filter
Photometric repeatability± 0.001 Abs (0.5 Abs) ± 0.001 Abs (1 Abs) ± 0.003 Abs (2 Abs) ± 0.1 %T
Noise level0.00003 Abs RMS (500 nm)0.00005 Abs RMS (500 nm)
Baseline flatness ± 0.0003 Abs (200–860 nm), 1 hour after light source is turned ON± 0.0004 Abs (200–860 nm), 1 hour after light source is turned ON
Baseline stability Within 0.0002 Abs/h (700 nm), 1 hour after light source is turned ONWithin 0.0003 Abs/h (700 nm), 1 hour after light source is turned ON
Sample compartmentInternal dimensions: W150 × D260 × H140 (mm) Distance between light beams: 100 mm Maximum optical path length of cell: 100 mm
DimensionsW450 x D600 x H250(mm)
Weight23 kg
Operating temperature15°C to 35°C
Operating humidity30 % to 80 % (no condensation, less than 70 % above 30 °C)
Power requirements100 to 240 VAC, 50/60 Hz
Power consumption170 VA 

 

Software Specifications UVProbe Software

Operating SystemWindows 7 Professional (64/32 bits)
Data Acquisition ModesSpectrum, Photometric (quantitation), and Kinetics (time course)
General
  • Multitasking (possible to execute data processing while measurement is being performed)
  • Customizable measurement screen layout (wavelengths, data display font and font size, colors, displayed number of digits)
  • GLP/GMP compliant (security, history)
  • Real-time concentration display
Spectrum Mode
  • Comparison/interactive processing of multiple spectra (Note 2)
  • Storage of all processed data with original data set including a history of all manipulations
  • Spectrum enlargement/shrinking and auto scale
  • Automatic interlink with spectral data processing
  • Annotation on spectrum screen
Data Processing in Spectrum Mode
  • Normalization, Point Pick, peak/valley detection, area calculation
  • 1st – 4th derivatives, smoothing, reciprocal, square root, natural log, Abs, %T conversion, exponential conversion, and Kubelka-Munk conversion
  • Ensemble averaging, interpolation, four arithmetic operations (between spectra, between spectra and factors)
Photometric (Quantitation) Mode
  • Single wavelength, multi-wavelength (includes 1, 2, or 3 wavelengths), spectrum quantitation (peak, maximum, minimum, area, etc. for specified wavelength ranges)
  • K-factor, single-point, multi-point calibration curves (1st, 2nd, 3rd order function fitting, pass-through-zero specification)
  • Photometric processing with user-defined functions (+, − , × , ÷ , Log, Exp, etc. functions, including factors)
  • Weight correction, dilution factor correction, and other corrections using factors
  • Averaging of repeat measurement data
  • Simultaneous display of standard sample table, unknown sample table, and calibration curves
  • Display of Pass/Fail indications
Kinetics (Time Course) Mode
  • Comparison/interactive processing of multiple time-course data (Note 2)
  • Single- or double-wavelength measurement (difference or ratio)
  • Simultaneous display of time-course data, enzyme table, and graphs
  • Enzyme kinetics calculation (for single- or multi-cell)
  • Michaelis-Menten calculations and graph creation (Michaelis-Menten, Lineweaver-Burk, Hanes, Woolf, Eadie-Hofstee), Dixon plot, Hill plot
  • Integrated management of sample information including original data, sample weight, and dilution factors, etc.
  • Event recording such as addition of reagents during measurement
  • Time-course spectrum data processing (same as in spectrum data processing)
Report Generator
  • Preview and print functions for customized formats
  • Layout and editing of templates
  • Quick printing using report templates
  • Auto-printing (spectrum mode)
  • Multi-page printout
  • Insertion of date, time, text, and drawing objects including lines, circles and rectangles
  • Insertion of spectral/quantitative data, method, and history
  • Insertion of headers and footers
  • Specification of graph line thickness (for each module), font style, and size