AP Series

Enquire Now

    W Series Analytical Balances 

    Series W Series
    Model AP225W AP135W AP125WD AP225WD AP124W AP224W AP324W
    Capacity 220g 135g 120g / 52g 220g / 102g 120g 220 g 320 g
    Minimum Display 0.01mg 0.01 mg / 0.1 mg 0.1 mg
    Calibration Weight Built-in *1
    External Calibration Weight Range for Span 95 to 220.00090 g 45 to 135.00090 g 45 to 120.00090 g 95 to 220.00090 g 45 to 120.00090 g 95 to 220.00090 g 95 to 320.00090 g
    Calibration (200g) (100 g)  (100 g) (200 g)  (100 g) (200 g) (300 g)
    Repeatability (Standard deviation) 0.015mg (to 20g) 0.03mg (to 100g) 0.05mg( to weighing capacity) 0.05mg 0.1 mg / 0.02 mg 0.1 mg / 0.05 mg 0.1 mg  0.15 mg
    Repeatability (for Low Loads) 0.01 mg (5 g low loads) 0.015 mg 0.1 mg (5 g low loads) 0.1 mg (10 g low loads) 0.1 mg (20 g low loads)
    Minimum Weight *2 20mg 30mg 200 mg
    Linearity *2 ±0.1mg ±0.1mg ±0.2 mg /±0.05 mg ±0.2 mg /±0.1 mg ±0.2 mg ±0.3 mg
    Response Time for Trace Measurements *3 2 sec.
    Response Time *4 8 sec. 2 sec. / 8  sec. 2 sec.
    USB Host (Type A) Included
    USB Device (Type B) Included
    Recipe Compounding Included
    HPLC Buffer Solution Preparation Included
    mol Conversion Function Included
    Sample (Concentration) Preparation Included
    Inspection Support Function Included
    Clock-CAL Included
    Ionizer Optional
    Operating Temperature/Humidity Range 5 to 40°C    20 to 85% *5
    Sensitivity Stability Against Temperature Range ±2 ppm/°C (10 to 30°C)
    Pan Size ø91 mm
    Body Dimensions Approx. 212 (W) × 411 (D) × 345 (H) mm (power supply unit included) Approx. 212 (W) × 367 (D) × 345 (H) mm
    Weight Approx. 7.9kg Approx. 7.0kg
    Display OEL display (dot matrix)
    Input/Output Terminal RS-232C (D-sub9P plug)    USB host (Type A)    USB device (Type B)    Ionizer

     

    X Series / Y Series Analytical Balances

     

    Series
    X Series  Y Series
    Model AP124X AP224X AP324X AP124Y AP224Y AP324Y
    Capacity 120g 220g 320g 120g 220g 320g
    Minimum Display 0.01mg
    Calibration Weight Built-in *1
    External Calibration Weight Range for Span Calibration 45 to 120.009 g (100 g) 95 to 220.009 g (200 g) 95 to 320.009 g (200 g) 45 to 120.009 g (100 g) 95 to 220.009 g (200 g) 95 to 320.009 g (200 g)
    Repeatability (Standard deviation) 0.1 mg  0.15 mg 0.1 mg  0.15 mg
    Repeatability (for Low Loads) 0.1 mg (5 g low loads) 0.1 mg (10 g low loads) 0.1 mg (20 g low loads) 0.1 mg (5 g low loads) 0.1 mg (10 g low loads) 0.1 mg (20 g low loads)
    Minimum Weight *2 200 mg
    Linearity *2 ±0.2 mg  ±0.3 mg ±0.2 mg  ±0.3 mg
    Response Time for Trace Measurements *3 2 sec.
    Response Time *4 2 sec.
    USB Host (Type A) Not Included
    USB Device (Type B) Included
    Recipe Compounding Not Included
    HPLC Buffer Solution Preparation Not Included
    mol Conversion Function Included Not Included
    Sample (Concentration) Preparation Included
    Inspection Support Function Included Not Included
    Clock-CAL Included Not Included
    Ionizer Optional Not Included
    Operating Temperature/Humidity Range 5 to 40°C    20 to 85% *5
    Sensitivity Stability Against Temperature Range ±2 ppm/°C (10 to 30°C)
    Pan Size ø91 mm
    Body Dimensions Approx. 212 (W) × 367 (D) × 345 (H) mm
    Weight Approx. 7.0 kg  Approx. 6.5 kg
    Display OEL display (dot matrix)
    Input/Output Terminal RS-232C (D-sub9P plug)    USB device (Type B)    Ionizer  RS-232C (D-sub9P plug)    USB device (Type B)
    1.  Minimum display 0.01mg models provide two internal weights as standard (see page12 for details).
    2. Be compliant with USP Chapter 41. This is the tested value by the weight of the balance’s capacity of 5% (or 5 grams’ weight). In the case of the AP225W, the results are for tests carried out with the optional internal windbreak plate applied. The minimum weight value is affected by the installation environment, so it is necessary to measure it in the actual environment of use.
    3. The response time for displaying 90% of added sample amount value in trace measurements (from 1mg)
    4. The response time value is typical.
    5. Non-condensing.