AUW-D Series

SeriesAUW-D (dual range semi-micro)AUWAUXAUY
ModelAUW220DAUW120DAUW 320AUW 220AUW 120AUX 320AUX 220AUX 120AUY 220AUY 120
Capacity220g /82g120g /42g320g220g120g320g220g120g220g120g
Minimum Display0.1 mg / 0.01 mg0.1 mg / 0.01 mg0.1 mg0.1 mg0.1 mg0.1 mg0.1 mg0.1 mg0.1 mg0.1 mg
Repeatability (Standard deviation,σ)≤0.1 mg (large range) ≤0.05 mg (small range)≤0.1 mg (large range) ≤0.02 mg (small range)≤0.15 mg≤0.1 mg≤0.1 mg≤0.15 mg≤0.1 mg≤0.1 mg≤0.1 mg≤0.1 mg
Linearity±0.2 mg (large range)±0.1 mg (small range)±0.2 mg (large range)±0.3 mg (small range)±0.3 mg±0.2 mg±0.2 mg±0.3 mg±0.3 mg±0.2 mg±0.2 mg±0.2 mg
Response time (typical)3 s (large range) 15 s (small range)3 s (large range) 12 s (small range)3 s
Operating ambient temperature5 to 40 °C
Temperature coefficient of sensitivity T T (10 to 30°C)±2 ppm/°C (When PSC is Off)±2 ppm/°C
Sensitivity stability against temperature change (When PSC is On, 10 to 30°C)±2 ppm*
Pan size80 mm dia approx.
Body dimensionsW220 mm × D330 mm × H310 mm approx.
Weight7 kg approx.
Power consumption7VA approx.
Backlighted display  ***     
Built-in  calibration weight********  
PSC********  
Clock-CAL********  
Inspection Support Function********  
Clock-CAL********  
Built-in clock********  
GLP/GMP/ISO calibration report********  
WindowsDirect**********
Formulation mode**********
Interval timer output**********
RS-232 I/F**********
Specific gravity measurement software, piece counting, % display**********
Analog display**********



  1.  Minimum display 0.01mg models provide two internal weights as standard (see page12 for details).
  2. Be compliant with USP Chapter 41. This is the tested value by the weight of the balance’s capacity of 5% (or 5 grams’ weight). In the case of the AP225W, the results are for tests carried out with the optional internal windbreak plate applied. The minimum weight value is affected by the installation environment, so it is necessary to measure it in the actual environment of use.
  3. The response time for displaying 90% of added sample amount value in trace measurements (from 1mg)
  4. The response time value is typical.
  5. Non-condensing.