LABTOUCH - AW
Ideal for laboratory and production
aW measuring instrument for:
- many samples, regardless of whether liquid, gel-like or solid
- partial temperature regulation, as well as integrated sample surface temperature measurement, enable accurate and reproducible measurements
- different measuring modes offer the user the possibility to carry out either a fast measurement on the production line or an exact AW determination for quality assurance of the products
- ergonomic design and touch screen makes water activity measurement easier than ever before
Meter Specifications LabTouch-aw Versions CM-2 Sensor/ CM-3 Sensor Measurement Specifications parameter Water activity Temperature Measuring principle Electrolytic Surface IR and NTC Measuring range 0.030…1.000aw 5…45°C (41…113°F) Adjustment range 0.110…0.970aw N/A Resolution 0.001aw 0.1°C Accuracy +/-0.005aw within the calibration range +/-0.10°C Precision +/-0.003aw within the adjustment range* +/-0.1°C Repeatability +/-0.002aw within calibration range* N/A Temperature control Programmable measuring temperature in the range of 15…30°C (59…86°F)* Adjustment point (%rH) 11%, 33%, 53%, 58%, 75%, 84%, 90%, 97% Mains connection 5 VDC, +/-6%, max. Power consumption 10W Power supply 90…264 VAC 50/60 Hz Display 4.3“ LC-colour display Communication > USB for PC / printer (Advanced) > SD/SDHC-card (Basic and Advanced) Housing > Two-part PC/ABS housing, weight: 2.8kg > Dimensions (L x W x H) = 200x285x105mm Measuring chamber volume 15mL Dimensions sample container Diameter 30mm x height 14mm Protection class IP22 21CFR 11 partially compliant