FEATURES
- A large sample pan allows even a large amount of sample to be placed evenly in a thin layer. The result is accurate and fast measurements.
- Mid-wave infrared quartz heater provides effective drying without interference for a wide range of samples. Besides the excellent drying performance, it offers a long operational life of 20,000 to 30,000 hours.
- The internal precision weighing balance is engineered with a Shimadzu UniBloc cell. The mechanism provides excellent stability and a long operational life against repeated temperature changes.
- Digital control allows a selection of measurement modes. 10 sets of measurement settings can be stored for quick recall. Select one of the 9 combinations of drying and halting modes to optimize the measurement of your sample.
- The weight loss rate in the previous thirty seconds is monitored and visually presented in the bar graph display. This feature is especially useful to show that the measurement is close to completion.
- Shimadzu’s unique WindowsDirect function is standard. Measurement data can be sent to an application such as Excel® without any software installation to the computer. All you need to add is an RS-232C cable. If you'd like to use "WindowsDirect" with "Windows 7", "Windows Vista", or a USB port, please contact our distributors.
- A larger sample pan contributes to accurate measurements, but the larger heat capacity normally produces larger zero drift due to temperature fluctuation. The MOC-120H is equipped with a unique auto-taring mechanism, which adjusts the zero drift automatically and ensures high accuracy, even with a larger sample pan.
- Bias function allows adjustment to the data obtained by other measuring methods or other testers.
- Large backlit LCD is easily read even under poor lighting conditions.
Model (P/N) | MOC-120H (321-63300-10 |
Measuring method | Heat drying and weight loss |
Sample pan size | 130 mm dia |
Minimum display in weighing | 0.001 g |
Measurement range of moisture content | 0.01% to 100.00% |
Moisture content minimum display | 0.01% |
Sample capacity | 120 g |
Measurement modes | Automatic or Timed ending modes, Standard, Rapid, Slow and Step drying modes, Predictive Measuring mode |
Drying heater | Mid-wave infrared quartz heater |
Setting temperature range | 30 to 200°C by 1°C increments (Sample position temperature) |
Dimensions | 220W × 415D × 190H (mm) |
Weight | 4.5 kg |
Operational temperature and humidity range | 5 to 40°C, 85% RH or lower |
Power requirements | AC 100 to 120 / 220 to 240 V, 640 W maximum |
Standard accessories | Sample pan 2 pcs, Sample pan handler 2 pcs, Aluminum sheet 20 pcs, Spoon, Spatula |
Stored procedures | 10 |