Model CB-S 170

Enquire Now

    Benefits

    • Maximum usable space with minimal footprint and excellent ease of use
    • Minimum risk of contamination thanks to uncompromising hot air sterilization
    • Fanless, easy-to-clean hygienic interior without fixtures
    • Simple and fast cleaning thanks to a seamless inner chamber

    Important characteristics

    • Temperature range: +6 °C above ambient temperature to +50 °C
    • Humidity range: 90 to 95% RH
    • CO₂ control range: 0 to 20 vol.% CO₂
    • The tried-and-tested anti-contamination concept with 180 °C hot air sterilization
    • Humidification system with condensation protection
    • CO₂ gas mixing jet with Venturi effect
    • CO₂ sensor with infrared technology
    • LCD to display temperature and CO₂ along with additional information and alarms
    • Internal data logger, measured values can be read out in open format via USB
    • Tightly fitted inner door made of safety glass
    • Seamless stainless steel inner chamber
    • Perforated shelves, stainless steel
    • Troubleshooting system with visual and acoustic alarms
    • Interfaces: Ethernet, USB, zero-voltage alarm contact
    • Units can be stacked with stacking adapters; the CB-S is suitable for stacking with CB and CBF models of the same size

    Optional equipment

    • Copper interior fittings consisting of copper shelves and copper-plated water pan
    • Stacking adapter for secure, space-saving installation
    • CO₂ gas tank changer for connecting two gas tanks to one or two incubators
    • CO₂ shaker with corrosion-resistant components in a stainless-steel housing
    • APT-COM Multi Management Software – for managing, recording and documenting unit parameters
    • Services – wide range of services to ensure that the unit functions correctly
    Designation CBS170UL-120V¹ CBS170-230V¹
    Option model Standard Standard
    Order number 9040-0190 9040-0189
    Performance Data Temperature
    Recovery time after 30 seconds door open at 37 °C [min] 6 6
    Temperature fluctuation at 37 °C [± K] 0.1 0.1
    Temperature range +6 °C above ambient temperature to 50 °C 50 °C
    Temperature variation at 37 °C [± K] 0.3 0.3
    Performance Data Climate
    Humidity range [% RH] 90…95 90…95
    CO₂-Data
    CO₂ measurement IR IR
    CO₂ range [Vol.-% CO₂] 0…20 0…20
    CO₂ Recovery time after 30 seconds door open at 5 Vol.-% CO₂ [min] 5 5
    Electrical data
    Nominal power [kW] 1.3 1.3
    Phase (Nominal voltage) 1~ 1~
    Power frequency [Hz] 50/60 50/60
    Rated Voltage [V] 100…120 200…230
    Unit fuse [A] 16 10
    Doors
    Inner doors 1 1
    Unit doors 1 1
    Housing dimensions, not incl. fittings and connections
    Depth net [in] 28.17 28.17
    Height net [in] 34.28 34.28
    Width net [in] 26.79 26.79
    Internal Dimensions
    Depth [in] 19.9 19.9
    Height [in] 23.64 23.64
    Width [in] 22.06 22.06
    Measures
    Interior volume [cu.ft.] 6 6
    Load per rack [lbs] 22 22
    Net weight of the unit (empty) [lbs] 198 198
    Permitted load [lbs] 66 66
    Wall clearance back [in] 3.94 3.94
    Wall clearance sidewise [in] 1.97 1.97
    Environment-specific data
    Energy consumption at 37 °C [Wh/h] 45 45
    Sound-pressure level [dB(A)] 41 41
    Fixtures
    The number of shelves (std./max.) 2/6 2/6