SENSO DIRECT - PH / CONDUCTIVITY / TDS / OXI / TEMP

    • Premium optical system with the reference beam
    • Automatic test recognition with internal barcode reader
    • Automatic cuvette type detection
    • Support of analytical quality assurance

    The XD series instruments are ideal for routine and spectral analysis. As true all-rounders, the XD 7000 and XD 7500 facilitate work routines, especially with automatic method recognition using barcodes, automatic cell recognition, and more than 150 pre-programmed methods. In addition to analytical quality control, they also support Good Labor Practice (GLP). The instruments are available together with bar-coded test tubes and a wide range of accessories from a single source - which is also easy on your wallet. High-quality reference beam optics, usability, easy handling, and multilingual global applicability round off the versatility of the VIS and UV/VIS spectrophotometers, which are also equipped for flexible on-site use.

    Applications

    • Boiler Water
    • Cooling Water
    • Disinfection Control
    • Drinking-Water Treatment
    • Galvanization
    • Pool Water Control
    • Pool Water Treatment
    • Raw Water Treatment
    • Waste Water Treatment
    OpticsGrid monochromator with a reference beam and beam splitter after exit slit
    MeasurementConcentration, single and multi-wavelength measurement of absorbance and % transmission, kinetics, spectra
    Wavelength Range320 - 1100 nm
    Wavelength Resolution1 nm
    Wavelength Accuracy± 1 nm on all Holmium peaks
    Wavelength Reproducibilitybetter than 0,5 nm
    Spectral Scope4 nm
    Photometric Range-3.3 - +3.3 Abs
    Photometric ResolutionAbsorption: 0.001 ; Transmission: 0.1 %
    Photometric Accuracy0.003 Abs below 0.6 Abs ; 0.5 % from 0.6 to 2.0 Abs
    Photometric Reproducability0.003 Abs below 0.6 Abs ; 0.5 % from 0.6 to 2.0 Abs
    Photometric Linearity< 1 % up to 2.0 Abs between 340 to 900 nm
    Scan Speed700 - 2000 nm/min.
    Stray Light< 0.1 % Transmission at 340 and 408 nm
    Drift< 0.005 Abs per hour after 15 minutes heat uptime
    OperationMembrane Keyboard
    Display7“ high contrast color graphic-display
    Suitable VialsRectangular Cuvettes 10 mm
    Rectangular Cuvettes 20 mm
    Rectangular Cuvettes 50 mm
    Round Cuvettes 13 mm
    Round Cuvettes 16 mm
    Round Cuvettes 24 mm
    Automatic Cuvette RecognitionRound cuvettes: 13, 16 and 24 mm ; Rectangular cuvettes: 10, 20 and 50mm
    Test Recognitionvia internal barcode reader
    InterfacesEthernet
    USB B
    USB A for External Memory
    Keypad
    Barcode-Scanner
    PCL Compatible Printer
    Auto-CheckSelf-test at each switch-on: a test of memory, processor, an internal interface, filter lamp, and additional calibration of each wavelength
    LIMS CompatibilityASCII, .csv-files
    Internal Storageapprox. 5000 data sets (method, user ID, date, result), auto storage function / manual storage function
    SecurityPassword protection possible: 3 different user levels (guest, user, admin)
    Power Drain100 - 240 V, 50/60 Hz
    Power SupplyBuffer batteries (4 x AA), power supply unit with cable
    Auto – OFFYes
    PortabilityBenchtop
    Environmental Conditions+10 °C to 35 °C (41 °F to 95 °F),
    ≤ 75 % average humidity within one year 95 %, max. 30 days/year, 85 % all other days
    Stock Conditions-25 °C to +65 °C (-13 °F to 268 °F)
    ComplianceCE
    Protection ClassIP 30
    IP Protection ClassEN 60529
    Interference EmissionClass B
    Interference ImmunityIEC 61000-4-3
    Tolerance Extension0.008 E
    Meter SafetyEC Directive 2014/35/EC EN 61010-1:2010
    Languages User InterfaceGerman, English, French, Spanish, Italian, Portuguese, Polish, Indonesian, Russian, Chinese, Japanese, Dutch, Swedish, Norwegian, Czech, Romanian, Macedonian, Slovenian, Hungarian, Turkish, Korean, Vietnamese, Thai, Serbian, Malaysian, Danish, Bulgarian